Analytical Sciences


Abstract − Analytical Sciences, 31(7), 693 (2015).

Enzyme and Mediator-coadsorbed Carbon Felt Electrode for Electrochemical Detection of Glucose Covered with Polymer Layers Based on Layer-by-Layer Technique
Soichi YABUKI and Yoshiki HIRATA
Division of Biomedical Research, National Institute of Advanced Industrial Science and Technology (AIST), 1-1-1 Higashi, Tsukuba, Ibaraki 305-8566, Japan
Glucose dehydrogenase (GlDH) and ferrocene were coadsorbed on a carbon felt (CF) sheet (5 × 10 mm, 2 mm thickness), which was used to construct an electrode for the electrochemical detection of glucose. A potential of +0.3 V vs. Ag/AgCl was applied on the base CF, and the current was measured. After the addition of glucose, the current increased and reached a steady state within 50 s. The current response was proportional to the glucose concentration up to 20 μM, with a lower detection limit of 1 μM. The surface of the CF electrode was covered by layers of polystyrene sulfonate and poly-L-lysine using layer-by-layer technique. Again the current response was proportional to glucose concentration up to 20 μM, with a lower detection limit of 2 μM. The oxidation current owing to electrochemical interferents such as L-ascorbate and acetaminophen was 1/8 times of the current observed on the unprotected electrode. In addition, the protection imparted stability to the electrode. Our work demonstrates that a GlDH/ferrocene CF electrode, protected with polystyrene sulfonate and poly-L-lysine, could be used for the electrochemical detection of glucose.