Analytical Sciences


Abstract − Analytical Sciences, 26(7), 797 (2010).

Spectrophotometric Determination of Trace Phosphate Ions by Amplitude-Modulated Flow Analysis Coupled with Malachite Green Method
Takeshi UEMURA,* Takeshi OGUSU,** Masaki TAKEUCHI,*** and Hideji TANAKA***
*Graduate School of Pharmaceutical Sciences, Tokushima University, 1-78-1 Shomachi, Tokushima 770-8505, Japan
**Faculty of Pharmaceutical Sciences, Tokushima University, 1-78-1 Shomachi, Tokushima 770-8505, Japan
***Institute of Health Biosciences, Tokushima University, 1-78-1 Shomachi, Tokushima 770-8505, Japan
An amplitude-modulated flow analysis coupled with lock-in detection is proposed for the determination of trace phosphate. The flow rate FS of sample solution is varied in response to a periodic signal Vc of alternating waveform. A coloring reagent solution (ammonium molybdate + Malachite Green) is delivered at a constant flow rate FR. Under the constant total flow rate FT, both solutions are merged with a diluent (water). Phosphate ion in the sample reacts with molybdate and then with Malachite Green to form green ion pairs in acidic media. Downstream, the absorbance of the mixed solution is measured at 625 nm. The output voltage Vd from the detector is sent to a lock-in amplifier, where the wave component of Vd that has the same frequency as that of Vc is distinguished from background signals. Phosphate ion can be determined from the amplitude of the component thus extracted. The calibration curve is linear (r2 > 0.998) and the limit of detection (3.3σ) is 0.17 μmol dm−3. Compared with a conventional flow-based method with no modulation, the present method is less susceptible to the baseline drift due mainly to the adsorption of the ion-pair on the optical window, because the quantification is based not on the absorbance itself but on the amplitude of the absorbance. Good recoveries around 100% are obtained for the phosphate ions spiked into real water samples.