Analytical Sciences


Abstract − Analytical Sciences, 20(7), 1025 (2004).

Determination of Trace Levels of Iron in a Seawater Sample Using Isotope Dilution/Inductively Coupled Plasma Mass Spectrometry
Yuichi TAKAKU,* Takuma HAYASHI,** Ikuko OTA,** Hidenao HASEGAWA,* and Shinji UEDA*
*Institute for Environmental Sciences, Department of Radioecology, 1-7 Ienomae, Obuchi, Rokkasho-mura, Kamikita-gun, Aomori 039-3212, Japan
**Tohoku Nuclear LTD. Advanced Technology of Science, 2-41-14 Higashi Okamisawa Misawa, Aomori 033-0024, Japan
An analytical method for trace levels of iron in a seawater sample using isotope dilution ICP-MS was developed. Preconcentration of iron and the removal of major elements in seawater such as alkali and alkaline-earth elements can be carried out quickly using a chelating resin disk by adjusting the sample pH to 3. The collision cell option of the ICP-MS instrument method was used to improve the performance of the instrument for iron measurements since ArO and ArN interferences could be reduced using this analytical method. About 4 ml min-1 helium, as the collision gas, were introduced into the cell. 40Ar14N and 40Ar16O which interfere with 54Fe and 56Fe in water had their amounts decreased by 5 orders of magnitude. Then, the isotope dilution method was used for iron determination below ng g-1 level of trace iron in four environmental reference materials (river water standard sample JAC-0031 (Japan Soc. for Analytical Chemistry), estuarine standard sample SLEW-2 (NRC Canada) and seawater standard samples CASS-3 and NASS-5 (NRC Canada)) were measured. Good agreement between analytical results and certified values of reference materials was obtained, which confirmed the effectiveness of this method.