A Sensitive Method for Cadmium Determination Using an On-line Polyurethane Foam Preconcentration System and Thermospray Flame Furnace Atomic Absorption Spectrometry
César Ricardo Teixeira TARLEY and Marco Aurélio Zezzi ARRUDA
Institute of Chemistry, Department of Analytical Chemistry, State University of Campinas-UNICAMP, P. O. Box 6154, Campinas SP 13084-971, Brazil
A new sensitive and low cost method for cadmium determination at µg l-1 levels that combines an on-line preconcentration system with the thermospray flame furnace atomic absorption spectrometry technique (TS-FF-AAS) is described in this work. Cadmium is preconcentrated from an acidic medium (pH 2.0) by forming a complex with ammonium O,O-diethyldithiophosphate (DDTP), which is then adsorbed onto polyurethane foam (PUF). The elution step is performed by using 80% (v/v) ethanol. The effects of the chemical and flow variables associated with the preconcentration were studied, such as the pH of formation of the Cd-DDTP complex, the DDTP concentration, the preconcentration and elution flow rate and the mass of adsorbent. The present method was operated in volume-mode (2 ml) and provided a linear range from 0.4 to 15.0 µg l-1 with a sample throughput of 16 h-1. The increase of power detection related to TS-FF-AAS by coupling the preconcentration system was confirmed by the enhancement of sensitivity (ca. 5 times), when compared to the value for TS-FF-AAS alone, thus achieving a low detection limit (0.12 µg l-1). The accuracy of the method was confirmed from analyses of spiked water samples and by the use of a reference technique (ETAAS). Certified biological materials were also used for the same purpose.