Analytical Sciences


Abstract − Analytical Sciences, 20(3), 471 (2004).

Flow-Injection Assay of Catalase Activity
Hiroyuki UKEDA, Yukiko ADACHI, and Masayoshi SAWAMURA
Department of Bioresources Science, Faculty of Agriculture, Kochi University, Monobe B-200, Nankoku 783-8502, Japan
A novel flow-injection assay (FIA) system with a double line for catalase activity was constructed in which an oxidase is immobilized and the substrate is continuously pumped to reduce the dissolved oxygen and to generate a given level of hydrogen peroxide. The catalase in a sample decomposed the hydrogen peroxide, and thus the increase in dissolved oxygen dependent on the activity was amperometrically monitored using a Clark-type oxygen electrode. Among the examined several oxidases, uricase was most suitable for the continuous formation of hydrogen peroxide from a consideration of the stability and the conversion efficiency. Under the optimum conditions, a linear calibration curve was obtained in the range from 21 to 210 units/mg and the reproducibility (CV) was better than 2% by 35 successive determinations of 210 units/ml catalase preparation. The sampling frequency was about 15 samples/h. The present FIA system was applicable to monitor the inactivation of catalase by glycation.