Analytical Sciences


Abstract − Analytical Sciences, 22(1), 99 (2006).

Micro Flow Injection Analysis Combined with a Separation Technique for the Urinary Glucose Assay
Qiong JIA,* Tomoko MATOBA,* Syouhei NISHIHAMA,** and Kazuharu YOSHIZUKA**
*Kitakyushu Foundation for the Advancement of Industry, Science and Technology, Hibikino 2-1, Kitakyushu 808-0135, Japan
**Department of Chemical Processes and Environments, Faculty of Environmental Engineering, The University of Kitakyushu, Hibikino 1-1, Kitakyushu 808-0135, Japan
A urinary glucose assay has been investigated, employing a micro flow injection analysis (μFIA) combined with a separation technique of glucose from the analyte. The adsorption part using activated alumina for the glucose in the analyte can be successively integrated onto a μFI chip. The selective adsorption-desorption of glucose in the artificial urine can progress on the adsorption part. Along with this selective preconcentration of glucose, the typical FI peak of glucose can be obtained just by feeding the sample and deionized water as an elutant sandwiched with the reagent on the carrier stream. The glucose concentration in artificial urine can be quantitatively determined with the present μFIA system, while the interference of other components coexisting in urine occurs in the case of the conventional FIA system without any separation part. The described method serves as a template for improving the selectivity for the analyte in the multi-component system.